ISO/ASTM 51538 covers the preparation, handling, testing and procedure for using the ethanol-chlorobenzene dosimetry system to measure absorbed doses in materials irradiated by photons and electrons in terms of absorbed dose in water. The system consists of a dosimeter and appropriate analytical instrumentation. This practice describes the titration analysis as a standard readout procedure for the ECB dosimeter and applies only to gamma rays, X rays and high-energy electrons.
This practice applies provided the following are satis-fied:
the absorbed dose range is from 10 Gy to 2 MGy;
the absorbed dose rate does not exceed 106 Gy s -1;
for radionuclide gamma-ray sources, the initial photon energy is greater than 0,6 MeV;
for bremsstrahlung photons, the initial energy of the electrons used to produce the bremsstrahlung photons is equal to or greater than 2 MeV;
for electron beams, the initial electron energy is equal to or greater than 4 MeV.
the irradiation temperature of the dosimeter is within the range from - 40 °C to 80°C;
the effects of size and shape of the irradiation vessel on the response of the dosimeter can adequately be taken into account by performing the appropriate calculations using cavity theory.
-
Текущий статус: ОтозваноДата публикации: 2002-03
-
Версия: 1
-
- ICS :
- 17.240 Radiation measurements
Жизненный цикл
-
Ранее
ОтозваноISO 15563:1998
-
Сейчас
-
00
Предварительная стадия
-
10
Стадия, связанная с внесением предложения
-
20
Подготовительная стадия
-
30
Стадия, связанная с подготовкой проекта комитета
-
40
Стадия, связанная с рассмотрением проекта международного стандарта
-
50
Стадия, на которой осуществляется принятие стандарта
-
60
Стадия, на которой осуществляется публикация
-
90
Стадия пересмотра
-
95
Стадия, на которой осуществляется отмена стандарта
-
00
-
Пересмотрен
ОтозваноISO/ASTM 51538:2009
Появились вопросы?
Ознакомьтесь с FAQ
Часы работы:
Понедельник – пятница: 09:00-12:00, 14:00-17:00 (UTC+1)