ISO 21647:2004
w
ISO 21647:2004
34637

Status : Withdrawn

This standard has been revised by ISO 80601-2-55:2011

Abstract

ISO 21647:2004 specifies particular requirements for the basic safety and essential performance of respiratory gas monitors (RGM) (as defined in 3.15) intended for continuous operation for use with humans. It supplements the requirements of IEC 60601-1:1988.

ISO 21647:2004 specifies requirements for anaesthetic gas monitoring, carbon dioxide monitoring and oxygen monitoring.

ISO 21647:2004 is not applicable to monitors intended for use with flammable anaesthetic agents.

The requirements of ISO 21647:2004 which replace or modify the requirements of IEC 60601-1:1988 and its Amendments 1 (1991) and 2 (1995) are intended to take precedence over the corresponding general requirements.

Environmental aspects are addressed in Annex CC. Additional aspects of environmental impact are addressed in ISO 14971.

General information

  •  : Withdrawn
     : 2004-11
    : Withdrawal of International Standard [95.99]
  •  : 1
     : 40
  • ISO/TC 121/SC 1
    11.040.10 
  • RSS updates

Got a question?

Check out our FAQs

Customer care
+41 22 749 08 88

Opening hours:
Monday to Friday - 09:00-12:00, 14:00-17:00 (UTC+1)