Abstract
This document specifies a method for calibrating the kinetic energy scales of Auger electron spectrometers with an uncertainty of 3 eV, for general analytical use in identifying elements at surfaces. This document also specifies a method for establishing a calibration schedule.
It is applicable to instruments used in either direct or differential mode, where the resolution is less than or equal to 0,5 % and the modulation amplitude for the differential mode, if used, is 2 eV peak-to-peak. It is applicable to those spectrometers equipped with an inert gas ion gun or other method for sample cleaning and with an electron gun capable of operating at 4 keV or higher beam energy.
General information
-
Status: PublishedPublication date: 2024-07Stage: International Standard published [60.60]
-
Edition: 3Number of pages: 11
-
Technical Committee :ISO/TC 201/SC 7ICS :71.040.40